*Automated gain, Peak Hold and Peak Memory;
*Automated display precise flaw location (Depth d, level p, distance s, amplitude,
sz dB);
*Ten independence setups, any criterion can be input freely, we can work in the scene without test block;
*Big memory of 300 A graph and 30000 thickness value.
*Gate and DAC alarm; Acoustic-Optical alarm;
*RS232 port, communication with pc is easy;
*The embeded software can be online updated;
*Li battery, continue working time up to 7 hours;
*Other assistant function;
*Display freeze;
*Automated echo degree;
*Angles and K-value;
*Lock and unlock function of system parameters;
*Dormancy and screen savers;
*Electronic clock calendar;
*Two gates setting and alarm indication;
*High-speed capture and very low noise;
*DAC, AVG; Solid metal housing (IP65);
*Provides high contrast viewing of the
waveform from bright, direct sunlight to complete darkness and easy to read
from all angles;
*Powerful pc software and reports can be exported to excel;
Designation |
Technical Data |
Range of scanning (mm) |
Range of scanning (mm):2.5~60000(Steel Longitudinal wave) Steps: 2.5,5,10,20, 30,40,50,60,70,80,90, 100,150,200, 250, 300, 350, 400, 450,500,600,700,800,900,1000,2000,3000,4000,5000,6000 Adjusting step: 0.1mm (2.5 mm~99.9mm),1mm(100mm~10000mm) |
D-delay (ms) |
D-delay (ms): -20~+3400 Steps: -20, -10,0.0, 10, 20, 50,100,150,200,250,300,350,400,450,500, 600, 700,800,900,1000,1500,2000,2500,3000,3400. Adjusting steps: 0.1(-20ms~999.9ms),1(1000ms~3400ms) |
P-delay (ms) |
P-delay:0.0~99.99 |
MTLVEL(m/s) |
MTLVEL:1000~9999 |
Working mode |
Single probe (receiving and sending), double probe (one for receiving and another for sending), transmission (transmission probe) |
Frequency Range (MHz) |
0.4–20 |
Gain adjustment (dB) |
0~110 Adjusting step: 0.0,0.2,0.5,1,2,6,12 |
Reject |
0%~80% of screen height, step: 1% |
Vertical linear error |
Vertical linear error is not more than 3% |
Horizontal linear error |
Not more than 0.2% in the scanning range |
Sensitivity Leavings |
≥62 dB |
Dynamic range |
≥32dB |
Alarm |
Three modes, i.e. forbidden wave, loss wave and auto |
Monitoring door |
2, expressed by bold transverse line, whose start, width and height are adjustable. Adjusting range of start (mm): horizontal pixel 0~208, the displayed value is relative with the scanning range. Step: value in mm corresponding to a pixel (relative with the scanning range) Adjusting range of width (mm): horizontal pixel 4~212, the displayed value is relative with the scanning range. Step: value in mm corresponding to a pixel (relative with the scanning range) Adjusting range of height: 2%~90% of vertical graduation Step graduation: 1% |
Display |
Display: high-brightness lattice |
A-Scan display area |
Full screen or local |
Data save |
300 A-Scan images (including setting of instrument) |
Standard communication interface with PC |
RS232 |
Measuring unit |
mm/inch |
Battery |
Li battery 4×3.6V4800mAh |
Power adaptor |
Input 100V~240V/50Hz~60Hz |
Working temperature |
-20℃~70℃ |
Working humidity |
20%~90% |
Port type |
BNC |
Dimension (mm) |
240×180×50 |
Weight (kg) |
1.8 |